LED Calibration

LED calibration is rig-specific. The goal is to map playlist intensity values to the optical output units used by the lab, then record the mapping in the rig configuration or analysis notes.

Operator Workflow

  1. Warm up the light source and measurement device according to local rig practice.

  2. Connect the DAQ output to the LED driver input used during experiments.

  3. Run a short playlist that steps through the expected intensity range.

  4. Measure optical output at the sample position.

  5. Update the rig’s attenuation/calibration notes.

  6. Re-run at least one validation intensity and record the measured value.

If the playlist uses MIRROR_LED, confirm that the DAQ protocol defines ledamp and that the generated pulse train has the expected voltage amplitude.